Maintenance contract Cleaning Line CR266

incl. 24/7 hotline service & 24-months warranty
 
brand new

The service and maintenance contract provides for a preventive maintenance interval of two years to ensure the best possible process safety and reliability of the CR266. Critical components, wearing parts and filters are replaced once a year or as required by the customer. We grant a 24-month warranty on all replaced components except wear parts. Telephone support is free of charge during this period.
Item no.: 37263
Add to favorites

Maintenance is carried out every two years by a CemeCon service technician on site at the customer's premises. Maintenance includes an inspection of the entire machine in accordance with the maintenance schedule:
 
  • Checking the switch cabinet-Checking the tank heaters
  • Replacement of pump head P8
  • Replacement of osmosis membrane and a pressure and flow check
  • Replacement of the spray arm
  • Replacement of all spray nozzles
  • Replacement of the chamber heaters and temperature sensors
  • Replacement of non-return valves on the working chamber
  • Checking the gear racks and gear wheels of the door drive
  • Check conductivity meter
  • Function test of all valves and replacement of all process ball valves
  • Replace pump P2 main pump
  • Internal inspection of all tanks, cleaning if necessary
  • Replacement of all tank seals
  • Replace door seal of working chamber
  • Check spray hoses; minor repair/replacement if necessary (max. 1 hour)
  • Replace filter, water inlet 2x, activated carbon, city water tank and demineralized water circuit
  • Replacement of the chamber ventilation air filter
  • Replacement of UV lamps for disinfection in VE water circuit
  • Checking the cooling system using pressure indicators in the operating theater and chart
  • Vacuum check; reading out the chart; manual check
  • Filling the system with water
  • Subsequent function check with test run
  • Filling the system with caustic solution and titration (if available at the customer's premises)
  • Preparation of a maintenance report
  • Discussion/advice on any further repair/replacement measures that may be necessary

If it is not yet necessary or not desired to replace the component, the components that have not been replaced will remain with you.